DESCRIPTION

UV-Vis Spectrophotometer SP-LUV1910, SP-LUV-1920:

1. Spectral bandwidth: The spectral bandwidth of the instrument is 1nm / 2nm, which ensures excellent spectral resolution and accuracy required for analysis.
2. Ultra-low stray light: excellent C-T monochromator optical system, advanced electronic system, to ensure ultra-low stray light level better than 0.03%, to meet the user’s measurement needs of high absorbance samples.
3. High-quality devices: The core devices are made of high-quality imported parts to ensure the stability and longevity of the instrument. For example, the core light source device is derived from the long-life deuterium lamp of Hamamatsu in Japan, which guarantees a working life of more than 2000 hours, greatly reducing the maintenance frequency and cost of daily replacement of the light source of the instrument.
4. Long-term stability and reliability: The design of the optical dual-beam optical system, coupled with real-time digital proportional feedback signal processing, effectively offsets the signal drift of light sources and other devices, ensuring the long-term stability of the instrument baseline.
5. High wavelength accuracy: The high-level wavelength scanning mechanical system ensures the accuracy of wavelengths better than 0.3nm and the repeatability of wavelengths better than 0.1nm. The instrument uses the built-in spectral characteristic wavelengths to automatically perform wavelength detection and correction to ensure long-term wavelength accuracy stability.
6. Light source replacement is convenient: the instrument can be replaced without removing the shell. The light source switching mirror supports the function of automatically finding the best position. The in-line deuterium tungsten lamp design does not require optical debugging when replacing the light source.
7. Instrument is rich in functions: The instrument is equipped with a 7-inch large-screen color touch LCD screen, which can perform wavelength scanning, time scanning, multi-wavelength analysis, quantitative analysis, etc., and supports the storage of methods and data files. View and print the map. Easy to use, flexible and efficient.
8. Powerful PC software: the instrument is connected to the computer via USB. The online software supports multiple functions such as wavelength scanning, time scanning, kinetic testing, quantitative analysis, multi-wavelength analysis, DNA / RNA analysis, instrument calibration, and performance verification. Supportauthority management, operation traceability, and meet various requirements in different analysis fields such as pharmaceutical companies.

SPECIFICATIONS

 
ModelSP-LUV1910/SP-LUV1920
Optical systemOptical double beam system
Monochromator systemCzerny-Turner monochromator
Grating1200 lines / mm high-quality holographic grating
Wavelength range190nm~1100nm
Spectral bandwidth1nm(SP-LUV1910) / 2nm(SP-LUV1920)
Wavelength accuracy±0.3nm
Wavelength reproducibility≤0.1nm
Photometric accuracy±0.002Abs(0~0.5Abs)、±0.004Abs(0.5~1.0Abs)、±0.3%T(0~100%T)
Photometric reproducibility≤0.001Abs(0~0.5Abs)、≤0.002Abs(0.5~1.0Abs)、≤0.1%T(0~100%T)
Stray light≤0.03%(220nm,NaI;360nm,NaNO2)
Noise≤0.1%T(100%T),≤0.05%T(0%T) ,≤±0.0005A/h(500nm,0Abs,2nm bandwidth)
Baseline flatness±0.0008A
Baseline noise±0.1%T
Baseline stability≤0.0005Abs/h
ModesT/A/Energy
Data range-0.00~200.0(%T)   -4.0~4.0(A)
Scan speedHigh / medium / low / very low
WL scan interval0.05/0.1/0.2/0.5/1/2 nm
Light sourceJapan Hamamatsu long-life deuterium lamp, imported long-life halogen tungsten lamp
DetectorPhotocell
Display7-inch large-screen color touch LCD screen
InterfaceUSB-A/USB-B
PowerAC90V~250V,  50H/ 60Hz
Dimension600×470×220mm
Weight18Kg
 