首頁 > 產品介紹 > infitek系列 > 體外診斷 > Blood Gas and Electrolyte Analyzer, BGA-95-8

產品分類

Blood Gas and Electrolyte Analyzer, BGA-95-8

  • Blood Gas and Electrolyte Analyzer, BGA-95-8

Description

Description

This new designed blood gas analyzer has a convenient disposable calibrants cartridge containing flush solution, mixed gas and liquid calibrants, waste container. This new cartridge design can avoid operator biological hazards and environment pollution.

Features

  • Intuitive touch screen

High resolution 10.4” touch screen

User-friendly operation interface

Parameter area provides analyzer status at a glance

Online operation and self-diagnostic tutorials

  • Infrared human-detector
  • Flexible and simple calibrants cartridge

High and low volume testing cartridge can be chose(300,200,150,100,50).

30 days available on board

Stored at room temperature

All-in-one design without extra consumables

Cartridge expiration alarm and residue information

  • Auto QC cartridge

Optional Auto QC cartridge containing 3 levels QC design

Auto H/M/L QC once per days, customer can define

30 days available on board

Stored at room temperature

All-in-one design without extra consumables

Cartridge expiration alarm and residue information

Auto-stop analysis sample once failing on QC

  • Reliable and convenient measurement unit

Capillary Sample volume only 50μL

Whole blood volume only 95μL

Multi-combination parameters

Self-made long life maintenance free electrodes

Maximum 90s result from aspiration, less than 120s from sample to sample

Long-lasting fluid system

Preheats reagents and sample

  • Easy to use

The sample adaptor is suitable for both syringe and capillary samples.

The sample probe’s wiper provides convenience, sample integrity andsafety

  • Fast data input

Easy sample and patient ID input

Convenient bar-code calibrants and Auto-QC cartridge replace

Easy bar-code consumables identification

  • Rich data management

RS-232 interface

LIS connection under HL7 protocol

Data backup with USB port

  • Compact, flexible

Easy move with rolling stand

Nickel-metal hydride battery,3.0Ah

Specifications

Measured Parameters

AnalyteMeasuring rangeUnit
Hydrogen ion activity(pH)6.000―9.000pH scale
Carbon dioxide0.667―26.67kPa
partial pressure(pCO2)5.0―200.0mmHg
Oxygen partial pressure(pO2)0―106.7

 

 

 

kPa
0―800.0

 

 

mmHg

 

 

Potassium ion(K+ )0.50―10.00mmol/L
Sodiumion(Na+ )50.0―200.0mmol/L
Chloride ion(Cl- )20.0―200.0

 

 

mmol/L
Ionized calcium(Ca++)0.10―5.00mmol/L
Hct12.0―65.0

 

 

%

 

 

B.P.500―800mmHg
Glucose1.1―66.7mmol/L
Lactate0.4―30.0mmol/L

Derived(Calculated) Parameters:

pH(TC)pH temperature corrected
pCO2 (TC)pCO2 temperature corrected
pO2 (TC)pO2 temperature corrected
TCO20-50mmol/L
HCO30-50mmol/L
Beb-25.0-25mmol/L
Beecf-25.0-25mmol/L
SBC0-50mmol/L
SO2%40-100.0%
Ca++(7.4)Ionized calcium“normalized” to pH 7.4
TCaTotal calcium content
AGAnion gap
RI0.0-70.0
AaDO20-700mmHg
O2SATOxygen Saturation
tHb(c)Hemoglobin(Calculated)
SampleWhole blood, Serum, Plasma, Dialysate, CSF
Test itemspH , pCO2, pO2, K+ , Na+, Cl– , Ca++, Hct
Input parameterPatient temperature

 

Hemoglobin

FiO

B.P.

InterfaceSerial Line RS-232

 

TCP/IP interface

USB port

Integrated Bar-code reader

Analysis time<90s
Sample volume95μL Whole blood

 

50uL Capillary blood

CalibrationAutomatic, programmable 1-point and 2-point calibration
Data storage>10000
Display10.4” TFT touch screen
BatteryNickel-metal hydride battery 3.0Ah
PrinterBuild-in Thermal printer70mm
Barcode readerUp to 14 kinds barcode type
Instrument(W*H*D)400*574*344mm
CAL PAK(W*H*D)258*180*70mm
AUTOQC(W*H*D)103*180*70mm
Weight30Kg
Work environment15℃~30℃
Storage environment5℃~40℃
Electricity100-240V~50/60Hz
 
TOP